The motor handpiece can stop automatically on detecting a low battery voltage.
After the motor handpiece is stopped for a short time, it may start again even
though the battery is drained. This is not a fault condition of the battery, it is
a normal characteristic.
The Lithium-Ion battery is recyclable, but disposal may not be permitted by law
in your location. Return them to your distributor if you are unsure of the proper
disposal method.
When disposing of the
X–SMART
Easy, follow the instructions of your local
government for disposal, as it may contain materials which could become
industrial waste.
When discarding the contra-angle attachment, dispose of it as medical waste.
Users are responsible for operational control, maintenance and inspection.
Store the system in a place where the temperature is at -10 to +60 ˚C (14 – 140 ˚F),
humidity at 10 – 85% RH, atmospheric pressure at 500 – 1060 hPa, and the system
is not subject to air with dust, sulfur or salinity.
Adverse Reactions
There are no known adverse reactions.
A. CLASSIFICATIONS OF EQUIPMENT
Type of protection against electric shock:
Class II equipment
Degree of protection against electric shock:
– Type B applied part
Method of sterilization or disinfection recommended by the manufacturer:
See Cleaning, Disinfection, & Sterilization
Degree of protection against ingress of water as detailed in the current edition of
IEC 529: IPX0
X–SMART
Easy complies to IEC60601-1 safety standard and the requirements of
the CE Marking of Conformity.