J. TROUBLESHOOTING
When trouble is found, check the following before contacting your distributor. If none
of these are applicable or the trouble is not remedied even after action has been
taken, the product may have failed. Contact your distributor.
X-SMART
EASY Charging Stand
Problem Cause Solution
5. Remove the new battery from the plastic bag.
6. Insert the battery connector into the receptacle
inside the battery chamber and place the
battery into the chamber with care so as
not to catch the cord.
PRECAUTIONS
In case of difficulty inserting the connector into the motor handpiece, make sure the connector
is properly oriented. Do not insert it by force.
7. Close the battery cover, put back the screw and plug. Ensure you charge the
battery before use.
PRECAUTIONS
The used lithium-ion battery is recyclable, but its disposal may not be permitted by law in your
location. Return them to your distributor if you are unsure of the proper disposal method.
Fig. 12
The power lamp does not turn
on when the charging stand is
plugged into the AC outlet.
Charging lamp light does not
light when handpiece is
inserted in charging stand.
The internal fuse has burnt.
No battery is inserted.
The charging contacts are
dirty.
The battery is defective.
Contact your distributor.
Insert battery.
Clean the contacts by rubbing
with a cotton swab.
Replace battery.
Charging indicator is blinking
Poor charging contact
connection
Defective batter
y
Defective charger
Reinsert handpiece into the
charging stand
Replace batter
y
Replace charging stand