TORQUE
The torque limit values can be adjusted.
Torque can be found in the top box.
If the torque limit is exceeded, the file will stop and “S.t” is shown in the torque
box on the display.
The present torque limit value is displayed. The
torque setting can be adjusted from 0.6 Ncm to
1.0 Ncm in increments of 0.2 Ncm, and from
1.0 Ncm to 4.0 Ncm in increments of 0.5 Ncm.
SPEED
The rotation speed can be adjusted.
The present rotation speed is displayed in the
middle box on the motor handpiece. The speed
setting can be adjusted from 100 to 800 rpm in
increments of 50 rpm.
AUTO REVERSE
The auto reverse mode can be selected. (See Auto Reverse Function, section 5.)
The default mode for auto reverse is ON. The last position used for auto reverse (on
or off) is what will be the new default until the user manually changes it. Only the
mode that is currently in use will be displayed.
ON If, during operation, the load reaches the
torque limit, the unit auto-reverses. If the
load is removed after auto reverse rotation,
it stops. See Auto Reverse Function, section 5.
OFF Auto reverse rotation is not activated.
BATTERY INDICATOR
The mark of the present remaining amount of the battery is displayed when the
handpiece is turned on.
Full charge
Approximately 30-80% remains
Less than 30% remains.
Batter
y is drained. Charge the battery.