This troubleshooting section is meant for use by qualified Cavitron
®
Service Technicians.

  
Intermittent scaling power or no
scaling power.
1. Insert malfunction.
2. Insert is not pushed in far
enough for automatic pick-up.
3. Malfunction in Steri-Mate™
Handpiece.
4. Bent or missing electrical pin in
Steri-Mate™ Handpiece.
5. Open or intermittent wires in
handpiece cable assembly.
6. Loose wiring or defective solder
joint in the unit wiring
7. Foot Control battery weak.
8. Foot Control not synchronized
to the unit.
1. Test with another Cavitron
®
insert. If test insert
works properly, discard the original insert.
2. a. Check if insert is fully seated in the handpiece.
b. If a handpiece soft grip is being used, verify
that the grip is flush with the hard plastic of the
insert port. Refer to the Installation
Instructions provided with the soft grip for
correct installation.
3. Replace Steri-Mate™ Handpiece.
4. Replace Steri-Mate™ Handpiece.
5. Install a working Steri-Mate™ Handpiece on the
cable. Unplug the Handpiece cable connector at J3
of the Power Drive PC Board and check the
continuity of the wires.
a. Connect the ohmmeter between RED-GRN wire
terminals. Flex the handpiece cable and check
for intermittent readings. If the ohmmeter
reading is not consistent or it is indicating an
open circuit, the handpiece cable assembly is
likely to be damaged and should be replaced.
b. Connect the ohmmeter between WHT-GRN wire
terminals and repeat the procedure above.
6. Troubleshoot the unit wiring and connectors.
7. a. Check information center for Low Battery
light indication. Replace battery as needed.
b. Connect an auxiliary Foot Control cable
between the Foot Control and the unit. The
unit can be operated with the cable until the
battery is replaced.
8. a. Follow Directions for Use & Service Manual
instructions for Foot Control synchronizing.
b. Connect the auxiliary Foot Control cable
between the Foot Control and unit. Unit can be
operated with the cable until the Foot Control is
synchronized.