Pressing anywhere on the top of the foot control
activates the system.



1. AC Power Cord
2. Dual Position Foot Control (Wireless)
3. Auxiliary Foot Control Cable
4. Cavitron 30K Ultrasonic Inserts
5. Cavitron DualSelect
Dispensing system
6. Cavitron Steri-Mate Sterilizable Handpiece
7. Cavitron Steri-Mate Grip (Available in a variety
of colors)

1. Cavitron Insert Replacement O-ring 12/Packs
PN: 62351 (black) for plastic and Bellissima inserts
PN: 62605 (green) for metal grips and prophy
2. Steri-Mate Handpiece Cable O-ring, PN: 79357
3. Lavage (Water) Filter, 10/Pack, PN: 90158
For detailed information, contact your local DENTSPLY
Professional Representative or authorized DENTSPLY
Professional Distributor.



The foot control is a two-positioned momentary switch.
The first position activates both the ultrasonic energy and
lavage at the insert tip. The second position activates the
Boost Mode. The Boost Mode (fully depressed foot control)
increases the ultrasonic power level for quick, efficient
removal of tenacious deposits without touching the system
base. To deactivate Boost Mode, release foot control to first
position.
NON-DEPRESSED
DEPRESSED
1st POSITION
DEPRESSED
2nd POSITION




Connect the Handpiece to the Cable Assembly by
aligning the electrical connections. If Cable Assembly
does not seat into the handpiece, gently rotate the
handpiece until contacts align, then fully insert handpiece.
Hold empty handpiece in an upright position over a
sink or drain. Activate the Foot Control until water exits
to bleed any air bubbles that might be trapped inside the
handpiece.
Lubricate the O-ring on the insert with water before
placing it into the handpiece. Fully seat insert with a
gentle push-twist motion. DO NOT FORCE.
Turn the Lavage Control to select flow rate during
system operation. Clockwise increases flow at insert tip,
counterclockwise decreases flow. The flow rate through
the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer
lavage. Higher flow rates produce cooler lavage. If
the handpiece becomes warm, increase the flow rate.
With experience the Dental Healthcare Professional will
be able to determine the best flow rate setting for
optimum operating efficiency and patient comfort.